Ultrasonic milk analyser

Operation manual

23.08.10

83/98

- Use fresh buffers for calibrations. Avoid contamination of the stock buffer

solution and do not use it beyond the expiry date.

- Keep all connections dry.
- Immerse the electrode far enough into the solution to insure the

reference junction is below the surface.

- Allow adequate time for the electrode to stabilize in standards and

samples before taking a reading.

- Clean the electrode periodically. Allow more time for aged electrodes.
- Do not use the pH electrode in solutions of fluoride ion at low pH. This

will etch the glass membrane.

- Sulphide vapors can permeate the electrode wick and contaminate the

reference element. Minimize contact in such environments and change
the reference electrolyte frequently.

Milk acidity and рН

In the table below the connection between pH value and acidity in

°T is

shown

°T

deviation

рН

average

value

°T

deviation

рН

average value

Raw milk

Pasteurized milk

16 6,74-6,70

6,72

16

6,68-6,64

6,66

17 6,69-6,65

6,68

17

6,63-6,58

6,61

18 6,64-6,58

6,62

18

6,57-6,52

6,55

19 6,57-6,52

6,55

19

6,51-6,46

6,49

20 6,51-6,46

6,49

20

6,45-6,40

6,43

21 6,45-6,40

6,43

21

6,39-6,35

6,37

22 6,39-6,35

6,37

22

6,34-6,30

6,32

23 6,34-6,30

6,32

23

6,29-6,24

6,26

24 6,29-6,24

6,25

24

6,23-6,19

6,21