LABEC M-MIA-CC User Manual

Page 66

Advertising
background image

Ultrasonic milk analyser

Operation manual

23.08.10

64/98


Rate = Fat * FatCo + SNF * SNFCo

(1)


Amount = Rate * Litres

(2)

Separately, for every measurement. Via option Prn Single rep the

operator may forbid or allow adding these results when, after each
measurement, is printing the rest of the measurement results on the serial
printer.

For all the measurements from the archive, the system calculates:

Average Fat = (Fat1*Litres1 + Fat2*Litres2…) / Litres1+Litres2…)(3)
Total Amount = Amount1 + Amount2 …

(4)

Average SNF = (SNF1*Litres1 + SNF2*Litres2…)/
(Litres1+Litres2…) (5)

Calculations are made for each separate calibration. Average results for Fat
and SNF may be used for control of total gathered milk.

In order to start working the report system needs the operator to enable it
(Report En/Dis), and to set the old report in initial adjustment (New Report).
Using IBM PC (via RS232 cable) user can enter the name of Milk collection
center and names of 2 working shifts (i.e. Morning shift 1,2).
During work the operator may delete the result from the archive for the last
measurement (if there is a need to be done). Before starting generating the
reports, coefficients, used for calculations have to be entered (Set Coeffs).
After the working shift is over, the operator has the possibility to print out
Short or Large Shift Report. User has possibilities to reject selected deliverer
number from report (Reject Deliv N).

9.9.3. Menus’ description:
Group Init:

Report En/Dis –enables/disables work of the system for archives and

reports

Report Enable

Yes

No OK Yes


Busy/Free Recs – shows the number of busy/free places in the archive.

Total number of records – 1000. When the analyser’s power supply is
switched on, the display shows the number of free records. During real

Advertising