2. The analyser is in mode Sleep – do not accept any commands, expects
change in the power (supply from switching adapter for starting the battery),
display shows the following


Low Battery



3. The battery power is under the allowed minimum, milk analyser is
automatically switched off. Before switching off, for a short time, the display
shows


Out of Battery


9.8. Automatic weighting the milk with scales (option).

Scales could be connected towards the milk analyser, which to be used for
weighting the delivered milk before measuring its parameters. For using the
scale the operator has to follow up the two procedures:

- Preparation for work and checking the scales
- Weighting delivered milk

9.8.1. Preparation for work and checking the scales
Put the scales on round and solid surface and level it, using its helical legs.
Checking the scales is done by using 2 standards of weight – light and heavy,
for example 5 and 50 kg, in mode:
Setup ->Accessories->Weigh Scale->Measure
(this is OFF Line measuring mode). In case there is discrepancy noticed, you
need to make scales calibration, as is described below.

9.8.2. Weighting delivered milk.
The procedure is analogous to work when entering data for the deliverer
using additional keypad. The difference is that weight of milk is automatically
send from the scales.
Consequence of work is as follows: switch on the analyser for normal work.
Place the vessel where the milk from different deliverers will be collected, for
example with 80 l volume. When the analyser is ready for work, the operator
has to press “#” (Enter) on the digital keyboard. Display shows the following: