1. Push the PGM Button once and then navigate using the

Up and

Down Arrows until you reac

h “CONF” in the bottom display.

2.

After you reach “CONF”, push the PGM Button again and use

Up

and Down Arrows to navigate until you reach “PRO” in the bottom
display and push the “PGM” button again.

3.

The display should now read “FNCT” in the bottom display and “2” in
the top display. You will need to change the “2” to “0”. To do that,
push the “PGM” button once and the “FNCT” should start flashing.
While it is flashing, use the

Down Arrow to change the “2” to “0”.

When the “FNCT” stops flashing, push the exit button repeatedly to
return to normal display (Actual Value on Top and Setpoint Value on
Bottom.)

4. Now the control is setup as a Single Setpoint Control. Just use the

Down Arrows to change Setpoint Value.

5. To use as programmable control, you must change back to

programmable control. To do this, in the “FNCT” display, change the
“2” back to “0”.

6.5 USING JUMO CONTROL AS A PROGRAMMABLE CONTROL.

1. The Jumo has a 24-step Set Point Time Base Programmable

Function with Power Exhaust Outlet ON or OFF. To setup a
program it first has to be entered in the control under program editor in

the PRO Section. To get to the PRO Section, push PGM Button

once and then the down arrow

once and the Bottom Display

should say PRO. To enter into the PRO Section, push the PGM

button

again and the display should say PrO1. This is the start

point of the first 8 steps. To start programming the first step, push the

PGM button

again and the lower display should say SP01. (01

indicates Step One) To move through the Step Menu, push the up or
down arrow

to move forward or backwards. The down arrow

moves you forward from Step 1 to 24 and the up arrow

moves you

backwards from Step 24 to 1. In each Step there are three categories,
SP, TP, and OC followed by a number. The number indicates the
Step Number.

SP = Set Point

TP = Time

OC = Contacts to be activated.

To change the value of SP, TP, and OC, push the PGM button.

The

SP, TP, OC will start flashing and then use up or down arrows

to

change the value. To program multiple steps, the step numbers must be
in sequence.