Unit will heat up at all

1/ verify that controller is asking for heat by looking for controller
light

– if controller light is not on, there is a problem with the

controller
2/check amperage

– amperage should be virtually at maximum rated

(data plate) amperage
3/ do all controller functions work?
4/ is the High Limit set high enough?

– for diagnostics, should be

fully clockwise with the pilot light never on
5/ has the fuse/circuit breaker blown?

Indicated chamber temperature
unstable

1/ ±0.1 may be normal
2/ is blower motor working?

– remove top panel and verify

movement of blower motor in shadow box
3/ is ambient room temperature radically changing

– either door

opening or room airflow from heaters or air conditioning ?

– stabilize

ambient conditions
4/ it may happen if exhaust stack is 100% open or if power exhaust is
cycling

– adjust stack to at least ¼ closed

5/ sensor miss-located, damaged or wires may be damaged - check
mounts for control and High Limit sensors, then trace wires between
sensors and controls
6/ calibration sensitivity

– call Customer Service

7 / high limit set too low

– be sure that High Limit is more than 5

degrees over desired set point; check if High Limit pilot is on
continuously; turn controller knob completely clockwise to see if
problem solved then follow instructions in Section 7.5 for correct
setting
8/ electrical noise

– remove nearby sources of RFI including motors,

arcing relays or radio transmitters
9/ bad connection on temperature sensor or faulty sensor

– check

connectors for continuity and mechanical soundness while watching
display for erratic behavior; check sensor and wiring for mechanical
damage
1 0 / bad connections or faulty solid state relay

– check connectors

for mechanical soundness and look for corrosion around terminals
or signs of arcing or other visible deterioration

Will not maintain set point

1 / assure that set point is at least 5 degrees over ambient room
temperature
2/ see if ambient is fluctuating

Display and actual reference
thermometer don’t match

1/ calibration error

– see Watlow users manual

2/ temperature sensor failure

– call Customer Service

3/ controller failure

– call Customer Service

4/ allow at least two hours to stabilize
5/ verify that reference thermometer is certified
See above

Can't adjust set points or calibration

1/ turn entire unit off and on to reset
2/ if repeatedly happens, call Customer Service

Calibrated at one temperature, but not
at another

This can be a normal condition when operating temperature varies
widely. For maximum accuracy, calibration should be done at or as
close to the set point temperature.

MECHANICAL

Door not sealing

1/ check physical condition of gasket
2/ assure that gasket is in original location