Troubleshooting, Temperature – Shellab SMO10HP-2 User Manual

Page 14

Advertising
background image

14

TROUBLESHOOTING

TEMPERATURE

Temperature too high

1/ controller set too high-see your Watlow users manual
2/ controller failed on

– call Customer Service

3/ wiring error

– call Customer Service

Display reads "HI" or "400"+

Probe is unplugged, is broken or wire to sensor is broken

– trace

wire from display to probe; move wire and watch display to see
intermittent problems

Chamber temp spikes over set point
and then settles to set point

Recalibrate

– see Watlow users manual

Temperature too low

1/ high limit set too low

– see section 4.5

2/ controller set too low

– see Watlow users manual

3/ unit not recovered from door opening

– wait for display to stop

changing
4/ unit not recovered from power failure or being turned off

incubators will need 24 hours to warm up and stabilize
5/ element failure

– call Customer Service

6/ controller failure

– confirm with front panel lights that controller is

calling for heat
7/ high limit failure

– confirm with front panel lights that High Limit is

operating correctly
8/ wiring problem

– check all functions and compare wiring to

owners manual - especially around any areas recently worked on
9/ loose connection

– check shadow box for loose connections

Display reads "LO"

1/ sensor is plugged in backwards

– call Customer Service

2/ if ambient room temperature is lower than range of unit

compare set points and ambient temperature to rated specifications
in Section 10.0

Unit will not heat over a temperature
that is below set point

1/ confirm that blower motor is moving and that amperage and
voltage match data plate

– check blower motor motion in shadow

box and feel for air movement in chamber
2/ confirm that set point is set high enough

–turn High Limit

clockwise and see if High Limit light comes on
3 / check connections to sensor
4/ check calibration

– using independent reference thermometer,

follow instructions in Watlow users manual

Section

8