EXAMPLE PROGRAM

Let’s say you want to have the oven to ramp up to 100°C in 30 minutes
and control for 1 hour at 100°C. Then you want to ramp up to 150° in 30
Minutes and control for 2 hours. Then you want to cool down to start Set
Point with Power Exhaust Outlet activated in one hour. See Graph:

EXAMPLE PROGRAM CHART

20

30

40

50

60

70

80

90

100

110

120

130

140

150

160

30 MINUTES INTERVALS

T

EM

PE

R

A

T

U

R

E

25

100

100

150

150

150

150

25


You would program the control like this:
STEP1

SP01=25°

TPO01=00:30hr

OC01-0000

STEP2

SP02=100°

TPO02=01:00hr

OC02-0000

STEP3

SP03=100°

TPO03=00:30hr

OC03-0000

STEP4

SP04=150°

TPO04=02:00hr

OC04-0000

STEP5

SP05=150°

TPO05=01:00hr

OC05-0100

STEP6

SP06=25°

TPO06=00:00hr

OC06-0000

The OC parameter 0100 is the value that activates the Exhaust Outlet on
the back of the oven.

To activate the program that is entered first return to normal display by

pushing the Exit Button repeatedly.

After Normal Display is

reached, push the up arrow

once. The lower display should flash STrT

and a RAMP/PROGRAM symbol

will appear in the lower right side of

the display. This indicates the program is running. To stop the program,
push the up arrow

again and the RAMP/PROGRAM symbol

will

disappear and the program will end. After program ends, control returns
to previous Set Point (starting Set Point). This is explained in Section 6.5
Paragraph 4.

6.6 OFFSET CALIBRATION

This is used to match the process temperature to an independent
thermometer. The Offset Calibration is measured in one degree
increments. For example, if the process value reads 100°C and the
independent thermometer reads 98°C, the Offset Calibration would be set