OPERATION

All control functions are covered in the Jumo Manual. This is a simplified
version.

Unit must be plugged into correct rating power source. Check user supplied
power source rating against serial plate. On ovens that need to be hard
wired, must be done by a qualified electrical technician.

6.1 Push the power switch to the I/ON position, turn the Over Temperature

Thermostat to its maximum position clockwise using a flat-head screw
driver. The main temperature controller should be illuminated. The top
display is the Process Temperature and the bottom display is the Set
Point
.

6.2 AUTO TUNING JUMO CONTROL FOR OPTIMUM CONTROL

Even though the control was Auto Tuned at the factory, you might choose
to tune the control to your specific Set Point.

1. Push the down arrow once.

2. Once the down arrow has been pushed, use the up and down

arrow

to achieve desired set point.

3. After reaching desired set point, the display should blink once to enter

the value.

4. To Auto Tune, push the up and down arrow

simultaneously for 2

seconds until the Set Point Display flashes Tune.

5. After Tuning Cycle is complete, the bottom display will stop flashing

Tune and return to desired Set Point.

6.3 CHANGING SET POINT ON CONTROL WHEN SET UP AS

PROGRAMABLE CONTROLLER

1. To change Set Point, push the

Down Arrow once and then use the

Up and Down Arrow to change the Setpoint.

CAUTION: If the

Up Arrow is pushed first, it will start a program and

will not allow you to change the Setpoint. To return to NORMAL
DISPLAY, push the

Up Arrow again.

6.4 CHANGING CONTROL FUNCTION TO SINGLE SET POINT

CONTROL

Section