Adjusting the shaker movement – Shellab SSI10R-2 User Manual

Page 14

Advertising
background image

SHEL LAB SI9/SI9R Shaking Incubator Operational Manual

14

Copyright © 2007 Sheldon Manufacturing. All rights reserved.

All specifications and features for this product are subject to change without notice.

Total Sample

Weight

Number of Counterweights

0.5”DIA. ORBIT

0.75”DIA. ORBIT

1.0”DIA. ORBIT

1.5”DIA. ORBIT

Start with these weights

2

3

5

7

Add weight per load:

2.5 kg (5.5 lbs.)

1

1-2

2

3

5 kg (11 lbs.)

2

3

4

6

7.5 kg (16.5 lbs.)

3

4-5

6

9

10 kg (22 lbs.)

4

6

8

12

12.5 kg (27.5 lbs.)

5

7-8

10

15

15 kg (33 lbs.)

6

9

12

18

17.5 kg (38.5 lbs.)

7

10-11

14

21

20 kg (44 lbs.)

8

12

16

24

3. Replace the wing-nuts and sample tray.

Adjusting the Shaker Movement

The shaker movement can be adjusted to: gentle, moderate or vigorous shaking. The shaking movement you use
depends on the oxygenation needed and cell strength of your samples. (Requires optional shaker mechanism parts and
instructions) Call SHEL LAB customer service for assistance.

Unit comes with 5 counter w

eights installed for a 1.0” diameter orbit configuration with no load and an extra 16 counter

weights for loads up to 20 kg (44 lbs).

(Extra counter weights and optional shaker mechanism parts can be ordered from SHEL LAB customer service)

WARNING

Always disconnect the unit from its power supply before attempting this procedure. Serious injury can result if
the drive plate operates accidentally.