A Typical Decontamination Procedure

Following is decontamination procedure that will suit most situations.

WARNING

Regardless of which decontamination procedure you follow, always turn the unit off and disconnect the service
cord from its power supply.

Before you reattach the unit to its power supply, be sure all cleaners are evaporated and dry.

To decontaminate the unit:

1.

Unplug the unit from its power source.

2.

Remove all interior parts, including shelves and shelf clips.

3.

Remove all gaskets and hinges. Clean and disinfect all mounting grooves for the door gaskets.

4.

Clean and disinfect all rubber or plastic tubing, as well as the fan and fan housing.

5.

Clean and disinfect all access ports, shaft holes, electrical feed points and any other passages into the chamber.

6.

Wash and disinfect all interior surfaces.

7.

Let the chamber dry out fully before replacing the removed parts or reattaching the unit to a power supply.

Control Maintenance

The main temperature controller, over-temperature protection thermostat and main temperature probe do not require any
maint

enance. If the unit appears to be having trouble maintaining a temperature, see “Troubleshooting” on page 17.