SHEL LAB SI9/SI9R Shaking Incubator Operational Manual

19

Copyright © 2007 Sheldon Manufacturing. All rights reserved.

All specifications and features for this product are subject to change without notice.

page 22

What is the problem?

Possible Causes

To solve the problem...

Cannot adjust set points or
calibration.

This is a controller failure.

1. Turn entire unit off and then

on to reset the unit. This
may temporarily solve the
problem, but controller may
be faulty.

If this does not solve the
problem...

2. Call customer service. See

page 22

Power

What is the problem?

Possible Causes

To solve the problem...

The unit will not turn on.

Power

cord

not

firmly

plugged into the outlet.

The unit or wall fuse/circuit
breaker has blown.

The outlet is defective.

The unit is plugged into a
circuit that already has too
many electrical loads.

1. Be sure the voltage and

frequency specifications of
the outlet are within the
range stated on the power
rating overlay at the rear of
the unit.

If this does not solve the
problem...

2. Check the power cord at the

electrical outlet for proper fit.

3. Make sure the unit is

plugged in firmly.

If this does not solve the
problem...

4. Replace fuse/circuit breaker

in the unit or wall as
necessary.

If this does not solve the
problem...

5. Make sure the outlet is in

proper working condition.

6. Replace if defective.

If this does not solve the
problem...

7. Check to see what other

loads are on the same circuit
as the unit. We recommend
that you plug your incubator
into a circuit separate from
other equipment.

8. Call customer service. See

page 22

The unit fuse/circuit breaker blows
often.

Wrong fuse installed.

Wire is shorting out.

1. Check fuse for right

amperage.

2. Call customer service. See

page 22