SHEL LAB SI9/SI9R Shaking Incubator Operational Manual

8

Copyright © 2007 Sheldon Manufacturing. All rights reserved.

All specifications and features for this product are subject to change without notice.

Installing the Unit

CAUTION

THIS INCUBATOR IS VERY HEAVY. DO NOT PERFORM THESE INSTALLATION

ADJUSTMENTS WITHOUT PROPER EQUIPMENT OR ADEQUATE ASSISTANCE.

Installation Overview

To install your unit, you need to:

1.

Select a suitable operating location for the unit.

2.

Level the unit.

3.

Sterilize the unit.

4.

Install the sample tray.

5.

Plug the unit into a power source.

Selecting a Location for the Unit

The operating location of you

r unit has a significant impact on your unit’s performance and how often it must be cleaned

and disinfected. Use the following guidelines to select the best location for your unit.

CAUTION

DO NOT MOUNT YOUR UNIT TO A FLAMMABLE SURFACE.

Operating Conditions For optimum performance, use your incubator at room temperatures between 18 and 25

C

(65 to 77

F) and at no greater than 80% relative humidity.

If you intend to operate the unit in conditions outside of these limits, contact your customer service representative.

Exposure. Avoid exposing the unit to the following:

Direct sun

High air movement, such as air vents, heating and cooling ducts, doors and other heavy traffic areas.

Extreme heat from steam radiators, stoves, ovens, autoclaves, or other sources of heat.

Level Surface. The unit must be located on a solid, flat and level surface strong enough to support its total weight.

Space requirements Allow a minimum of 20 cm (8 in.) between the rear and sides of the unit, and any walls or
partitions that can obstruct free airflow. Do not block access to the power cord, circuit breaker or fuses.

Cleanliness. Good laboratory quality control practice is the most efficient and reliable way to keep your unit free
from contamination.

If it is important that the interior of your unit remain sterile, always pay attention to the following guidelines:

Keep the air in the laboratory as clean as possible.

Keep the floor around the unit clean.

If the unit must be placed at floor level, use a sturdy platform. (See Option Accessory List Page 23)

Minimize the number of times access is made to the chamber during normal operation. Each time the door is
opened; room air is drawn in and can lead to contamination of the unit.

After deciding on the location for your unit, follow the installation instructions below.