ThermoWorks 292-701 SAF-T-LOG PAPERLESS HACCP THERMOMETER Operating Instructions User Manual

Page 5

Advertising
background image

5

Selecting Items
In measurement mode, the Up () / Down () arrow buttons are
used to scroll through the ‘Checklist Items’ The current ‘item name’
will be displayed in the Alpha/numeric portion of the LCD display
By holding down the Up () or Down () buttons the scrolling
speed will increase The corresponding temperature limits (HI) and
(LO) are displayed in the upper left portion of the display For Yes/
No items, ‘Yes or No’ will be displayed ‘PASS’ will be displayed if

the temperature being measured is between the high and low limits ‘FAIL’ will flash if the temperature being
measured is outside the high and low limits

Record
For temperature items pressing the ‘Record’ button saves the current temperature reading, together with date
and time, user, item name that is currently displayed, checklist name, PASS/FAIL status and limits to memory
If the reading falls outside of the acceptable limits, the user is prompted to store a corrective action

For Yes/No items, pressing the ‘Rec’ button enters the Yes or No selection function in the lower display
Scroll to either ‘Yes’ or ‘No’ to store the desired value If ‘No’ is selected, the user is prompted to include a
corrective action from the list of Yes/No corrective actions that is stored for the current checklist The unit can
store up to 1,000 readings After reaching 1,000 readings, the unit will no longer save readings and display
‘Memory Full ’

Instrument Menu
To enter the Instrument Menu on the Saf-T-Log device, press and hold the Up () and Down () arrows
simultaneously Within the Instrument Menu the user can: view saved readings, change checklists, change
the current user, access the instrument info menu to view date, time, firmware version, instrument name,
Checklist Name, calibration info, and TRIM value; and change the Auto-Off setting