System Setup

Section 6


Use the arrow keys to select the desired autostable % (the smaller the %, the more
accurate your reading and the more time it will take to achieve). Press CONFIRM. The
following display will appear:

S e t u p

# o f r d g s ?





Press CONFIRM. The following display will appear:

S e t u p

# o f r d g s = 5





Use the arrow keys to select the # of readings which must pass while complying with the
% change entered previously. (The larger the #, the greater the accuracy and the longer it
will take.)

6.3 SETTING COMMUNICATION PARAMETERS

The Model 52 will default to communications values that are appropriate for most
printers and personal computers. The defaults are Baud rate: 9600; parity = none; word
length = 8; and stop bits = 1. We recommend that you try these default settings before
attempting to change them. If you must change communication parameters on the Model
52, do so before using the RS232 port. Before beginning this sequence you must know
the baud rate, word length, number of stop bits, and parity of the PC in use. To find this
information, consult the documentation that accompanies your printer or PC. The Model
52 supports all common variations of these parameters.

To change a communication parameter, turn the function switch to SYSTEM SETUP.
Press SKIP repeatedly until the following display appears:

S e t u p

c o m m u n i c a t i on ?





YSI, Incorporated