CODE ERROR

CAUSE

CORRECTION

E. 0

System error

Defective ROM

Return for service

E.00

Lost calibration

Defective RAM backup

Return for service

value

battery

E.01

Defective RAM

Defective RAM

Return for service

E .1

Open circuit in

Connector improperly

Check connection

Temp probe

installed

Intermittent connection

Repair

or

replace

in cable or plug

Faulty temperature sensor

Repair or replace

E .2

High background

Insufficient warm-up time

See Preparation for Use

Incorrect probe zeroing

See Zero Calibration

procedure

Probe needs servicing

See probe instructions

Probe malfunction

Repair or replace

E .3

Low

Insufficient electrolyte

See probe instructions

sensitivity

Contaminated electrodes

See probe instructions

or fouled membrane

Membrane too thick

Try another membrane

High resistance in probe

Return for evaluation

connection

E .4

Output too high

Membrane too thin

Try another membrane

Short circuit

Repair or replace

Electrodes need

Repair or replace

resurfacing

Internal leakage in probe

Repair or replace

or cable connector

M.XX

Incorrect Mode

Switch or circuit defect

Return for service

(x may be any

digit or any letter)

YSI, Incorporated

Model 52

26