Discussion of Measurement Errors

Section 9


Calculation

TYPE

DESCRIPTION

CALCULATION

ERROR(mg/L)

1a

instrument accuracy ±0.001 x 8.24 + 0.01

±0.02

2a

probe background

±0.01 x (1-8.24/9.07) x 7.26 ±0.01

2b

probe nonlinearity

±0.003 x 7.26

±0.02

2c

temp. compensation

±(24-20) x 0.002 x 7.26 ±0.06

3a

temp. uncertainty

±0.01 x 7.26

±0.07

3b

pressure

±0.017 x 7.26

±0.12

3c

altitude

±0.018 x 7.26

±0.13


Probable error can be determined by an r.m.s. calculation:

r.m.s. error = [.022 + .012 + .022 + .062 + .122 + .132 ]1/2 = ±0.19 mg/L