Appendix a general specifications – YSI 52 User Manual

Page 41

Advertising
background image

APPENDIX A

GENERAL SPECIFICATIONS

Oxygen Measurement

Ranges:

0.0 to 19.99 mg/L dissolved oxygen

0.0 to 199.9% air saturation

Accuracy:

±0.1% of saturation value, plus 1 Least Significant Digit for mg/L
readings, plus probe error
±0.1% of air saturation, plus 1 Least Significant Digit for % readings, plus
probe error

Resolution: 0.1% of the saturation value, expressed in mg/L, or 0.01 mg/L , whichever

is greater; 0.1% air saturation

Temperature Measurement

Range:

-5.0 to +45.0°C

Accuracy:

±0.1°C, plus probe error

Resolution: 0.1°C

Temperature Compensation

The mg/L mode is automatically temperature-compensated to an accuracy of ±1% of DO
readings between 0 and 5°C, and to an accuracy of ±0.6% of readings between 5 and
45°C.
The % air saturation mode is automatically temperature-compensated to an accuracy of
±0.5% of calibration values between 0 and 5°C, and to an accuracy of ±0.3% of values
between 5 and 45°C.

Salinity Compensation

Range:

0.0 to 40.0 ppt

Accuracy:

±.02 mg/L

Operating Environment

0 to 45°C, 10 to 90% humidity, non-condensing

Water Resistance

With the probe receptacle capped, every case opening is has a gasket to resist the entry of
water.

YSI, Incorporated

Model 52

37