Discussion of Measurement Errors

Section 9


2.

Probe nonlinearity error: ±0.3% of reading

3.

The variation from nominal response to sample temperature is ±0.2% of the DO
reading per degree C of the temperature difference between the temperature of the
sample and the temperature at which the probe was calibrated.

9.3 TYPE 3 ERRORS - CALIBRATION

1.

Sample temperature uncertainty error: ±1% of reading

This error is zero when calibrating in the % air saturation mode or when
calibrating to a Winkler titration sample.

2.

Error due to barometric pressure uncertainty of 13mm Hg: ±1.7% of reading

3.

Error due to altitude estimation uncertainty of 500 ft: ±1.8% of reading

9.4

ERROR CALCULATION EXAMPLE

The example given presumes that air calibration is used. If the Winkler titration
calibration method is used, type 3 errors are replaced by the uncertainty attributable to
the overall Winkler determination. This example is for an extreme combination of
conditions.

Calibration conditions:

method:

air calibration

temperature:

24°C

altitude:

600 feet

calibrated to:

8.24 mg/L


Measurement conditions:

temperature: 20°C
reading: 7.26 mg/L
mode: mg/L