Entering the density directly< ,

q

,' )

In the specific gravity measuring mode, press and hold the [MODE |key until

the density currently set ( unit: g / cm^, factory setting : 1 .OOOOg / cm^) is

displayed. Use the following keys to change the value.

J zinnnn

0

;I.£IUUU

RE-ZERO I key

RANGE key
PRINT key

CALI key

Changes the numerical value of the digit selected.

Selects the digit to change the value.

Saves the change, displays "

f n d

" and returns to the

specific gravity measuring mode.

Returns to the specific gravity measuring mode without

saving the change.

Note

The range to set the density is 0.0000-1.9999 g / cm^. (Displayed up to four

decimal places)

Measuring the density

In the density measurement, the balance displays the weight of the sample
in air, the weight in liquid and then the density.

Measuring the weight of the sample in air.

I'JZtiJC

IL J IJ 9

The processing indicator flashes with the unit" g " displayed.

□ Measuring the weight of the sample in liquid.

on

1 C r

JJJ III J 9

The processing indicator illuminates with the unit" g " displayed.

Displaying the density.

0

C rinrin

4UUUU

The processing indicator illuminates with no unit displayed.

To switch between the above three, use the [RANGE key

3.Specific graviiy(densiiy)measuremenf

page 52