Entering the density directly< ,
q
,' )
In the specific gravity measuring mode, press and hold the [MODE |key until
the density currently set ( unit: g / cm^, factory setting : 1 .OOOOg / cm^) is
displayed. Use the following keys to change the value.
J zinnnn
0
;I.£IUUU
RE-ZERO I key
RANGE key
PRINT key
CALI key
Changes the numerical value of the digit selected.
Selects the digit to change the value.
Saves the change, displays "
f n d
" and returns to the
specific gravity measuring mode.
Returns to the specific gravity measuring mode without
saving the change.
Note
□
The range to set the density is 0.0000-1.9999 g / cm^. (Displayed up to four
decimal places)
Measuring the density
In the density measurement, the balance displays the weight of the sample
in air, the weight in liquid and then the density.
□
Measuring the weight of the sample in air.
□
I'JZtiJC
IL J IJ 9
The processing indicator flashes with the unit" g " displayed.
□ Measuring the weight of the sample in liquid.
on
1 C r
JJJ III J 9
The processing indicator illuminates with the unit" g " displayed.
Displaying the density.
0
C rinrin
4UUUU
The processing indicator illuminates with no unit displayed.
To switch between the above three, use the [RANGE key
3.Specific graviiy(densiiy)measuremenf