Weighing pan Error

This is a warning that the weight value is too light.

Confirm that the weighing pan and the pan

support are properly installed.

Unit weight, 100% weight error

The unit weight of the sample is very light in the

counting mode, or the 100% sample is too light

in percent mode. The balance can not calculate
it.

Increase the unit weight or 100% weight.

ARA Zero error

The ARA (Automatic Response Adjustment)
can not be performed, because there is some­
thing on the pan.

Remove all matter from the pan. Press the\CAL
key to return to the weighing mode.

ARA Unstable error

The ARA (Automatic Response Adjustment)

can not be performed because of unstable
weighing value.

Correct the environment for the balance. Press

the

I

CAL

I

key to return to the weighing mode.

Full memory

The memory data has reached 200 items.

When data is deleted, new data can be stored.

Memory data error

The memory data is lost.

Clear all memory data.

Unit weight information

This is advice regarding the sample number

that is needed to set the unit weight. When the

unit weight is computed and the sample num­

ber is too few, the required number is displayed

for counting accuracy.

Count and place the samples on the pan. Press
the

[

print

]

key to store the correct value.

Other errors

if you can not cancel the error yourself, request service from the store where

you purchased the balance or option, or the A&D service group.

Maintenance

page 67

17, Care and error codes