Signal connection – ADLINK PXI-7901 User Manual

Page 15

Advertising
background image

Signal Connection • 9

3

Signal Connection

3.1 PXI-7901 Topology

The PXI-7901 has 16 independent SPDT Relays onboard. Each relay has three
terminals, NC (normally closed), NO (normally opened), and COM (common).
When a relay is in its reset (de-energized) state, the NC connects to COM;
while in set (energized) state, the NO connects to COM. Thus all signal paths
on PXI-7901 are inherently break-before-make.

After power-up, all relays are in their reset (de-energized) state; i.e. NC
connects to COM.

3.2 PXI-7901 Pin assignment & Description

22. COM0

43. NO0

23. COM1

1. NC0

44. NO1

24. COM2

2. NC1

45. NO2

25. COM3

3. NC2

46. NO3

26. COM4

4. NC3

47. NO4

27. COM5

5. NC4

48. NO5

28. COM6

6. NC5

49. NO6

29. COM7

7. NC6

50. NO7

30. COM8

8. NC7

51. NO8

31. COM9

9. NC8

52. NO9

32. COM10

10. NC9

53. NO10

33. COM11

11. NC10

54. NO11

34. COM12

12. NC11

55. NO12

35. COM13

13. NC12

56. NO13

36. COM14

14. NC13

57. NO14

37. COM15

15. NC14

58. NO15

38. NC

16. NC15

59. NC

39. NC

17. NC

60. AUX3

40. AUX4

18. AUX2/SHDNn

61. AUX6

41. +5V out

19. AUX5

62. AUX7

42. AUX1/TRG_IN 20. GND

21.

AUX0/S_ADV

Table 1: Pin Assignment