Standard, Auto system operational test – Dukane ULTRASONIC GENERATOR/POWER SUPPLY 403-573-02 User Manual

Page 35

Advertising
background image

Page  2

Page  2 

iQ Series,

 Auto User’s Manual

Section 4 –System Operation

Page  2

Page  2 

iQ Series,

 Auto User’s Manual

Section 4 –System Operation

Dukane Manual Part No. 403-573-02

Standard

iQ

Auto System

Operational Test

Step 1. Verify that the standard system installation is

complete and all cables are connected.

Refer to installation instructions included in Sec-
tion 3, if needed.

Refer to Section 6 - Automation Interface -for in-
formation on wiring system controls, if needed.

Step 2. After completing Step 1, activate line power to

the standard iQ Auto system.

Normal Condition:

GREEN - The POWER STATUS

LED

on the iQ

Auto panel should light up GREEN.

The system is now ready to operate.

Troubleshooting Abnormal iQ Auto System
Conditions

RED - If the POWER STATUS

LED

lights up

RED, check line voltage level.

GRAY - If the POWER STATUS

LED

is a

gray color (not lit), check line input.

Optional System Status Output to Monitor:

Optionally the Power OK status output can be
monitored on I/O connector Pin 6.

This NPN status output signal will activate when
power is OK.

Step 3. After completing Step 2, test ultrasound output

by activating I/O connector Pin 12.

Normal Condition:

The system is operating properly when power is
delivered to the attached stack.

Ultrasound Activation Connections

:

Usually an automated control system is wired to
this control input.

Optionally - A manual switch could be wired to
this control input.

Figure 4-1

Generator POWER STATUS

LED

Detail

Options for U/S Activation Connections:

Optionally, monitor the U/S Active status output on
I/O connector Pin 4.

This NPN status output signal will activate when U/S
is ON.

Step 4. Other Options the Operator can Check:

Fault Status output signals

Fault Status output signals are available for:

Overload (Pin 3), or

Any Fault (Pin 5).

These NPN status outputs will activate when there is
a fault:

An Overload Fault latches until the next time U/S

is activated, or Fault Reset Input is activated..

Any Fault status activates when any fault is de-

tected by the system. It latches until the start of the
next cycle, or Fault Reset Input is activated.

Continued