IEEE 488.2 Commands—Description

*CLS

Description

The *CLS command clears the Standard Event

Status Register and the Error/Event Queue.

Syntax

*CLS

Parameter(s)

None

*ESE

Description

The *ESE command sets the Standard Event

Status Enable Register bits, as defined in the

table below. This register contains a mask value

for the bits to be enabled in the Standard Event

Status Register.

Syntax

*ESE<wsp><RegisterValue>

Parameter(s)

RegisterValue:
The program data syntax for <RegisterValue> is

defined as a <DECIMAL NUMERIC PROGRAM

DATA> element.
The <RegisterValue>, expressed in base 2,

represents the bit values of the Standard Event

Status Enable Register.

PON

CME

EXE

DDE

QYE

N.U.

OPC

MSB

LSB

Standard Event Status Enable Register

N.U.

Advertising