Calculate[1..n]:rbw – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) User Manual
Page 477
Advertising

SCPI Command Reference
Optical Spectrum Analyzer
469
Product-Specific Commands—Description
:CALCulate[1..n]:RBW
Description
This command sets the RBW for OSNR. This
value is used in calculation only if RBW
Instrument is OFF (or 0).
For this command to be accepted, the OSA
module must be in Ready state with no
acquisition in progress.
At *RST, this value is 0.100 nm.
Syntax
:CALCulate[1..n]:RBW<wsp><RbwForOsnr>
Parameter(s)
RbwForOsnr:
The program data syntax for <RbwForOsnr> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
Selects the RBW for OSNR.
Range: [0.010 … 2.000] nm
Example(s)
CALC:RBW 0.1
See Also
CALCulate[1..n]:RBW?
CALCulate[1..n]:RBW:INSTrument
CALCulate[1..n]:RBW:INSTrument?
Advertising