Sense[1..n]:average:inband:count – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) User Manual
Page 506
Advertising

SCPI Command Reference
498
FTB-5240S/5240BP
Product-Specific Commands—Description
:SENSe[1..n]:AVERage:INBand:COUNt
Description
This command sets the number of In-Band scans
that will be averaged to produce final trace.
For this command to be accepted, the OSA
module must be in Ready state with no
acquisition in progress.
At *RST, this value is 64.
Syntax
:SENSe[1..n]:AVERage:INBand:COUNt<wsp>
<InbandCount>
Parameter(s)
InbandCount:
The program data syntax for <InbandCount> is
defined as a <numeric_value> element.
Selects the Number of scans used to inband for
the final scan.
Range: [4 ... 9999]
Example(s)
SENS:AVER:TYPE AVERAGING
SENS:AVER:INBand:COUN 4
SENS:AVER:COUN 12
INITiate:IMMediate
See Also
SENSe[1..n]:AVERage:INBand:COUNt?
SENSe[1..n]:AVERage:TYPE
Advertising