SCPI Command Reference

502

FTB-5240S/5240BP

Product-Specific Commands—Description

:SENSe[1..n]:WAVelength[:RANGe]:

STARt

Description

This command sets the start wavelength for

acquisition range.

For this command to be accepted, the OSA

module must be in Ready state with no

acquisition in progress.

At *RST, this value is set to 1530.0 nm (C-band

lower limit).

Syntax

:SENSe[1..n]:WAVelength[:RANGe]:STARt<wsp

><Start[<wsp>NM]>|MAXimum|MINimum

Parameter(s)

Start:
The program data syntax for <Start> is defined

as a <numeric_value> element followed by an

optional <SUFFIX PROGRAM DATA> element.

The allowed <SUFFIX PROGRAM DATA>

element is NM. The <Start> special forms

MINimum and MAXimum are accepted on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

Minimum wavelength for acquisition range.

Range is device-dependent.