Far-end – EXFO FTB-8100 Series Transport Blazer for FTB-200 User Manual
Page 379

Common Tabs
SONET/SDH Application
367
Performance Monitoring (PM)
Far-End
³
EFS (Error Free Second): Gives the count of the seconds within which
no error occurred or when a defect is detected on the near-end.
³
EC (Error Count) (G.821 only): Gives the number of bit errors.
³
EB (Errored Block) (G.826, G.828, and G.829): Gives the count of
blocks in which one or more bits are in error.
³
ES (Errored Second): For G.826, G.828, G.829, M.2100 ISM, and
M.2101: Gives the count of the seconds within which one or more
anomalies (FAS, EB, etc.) occurred or at least one defect occurred.
³
SES (Severely Errored Second):
For G.826, G.828, G.829, and M.2101: Gives the count of the seconds
within which anomalies (FAS, EB, etc.) is ≥ X% or at least one defect
occurred. X = 30% for DSn/PDH signals; see the following table for
SONET/SDH signals.
STS-1
STM-0
OC-3
STM-1
OC-12
STM-4
OC-48
STM-16
OC-192
STM-64
Path
30%
30%
30%
30%
30%
Line/MS
15%
15%
25%
30%
30%
Section/RS 10%
30%
30%
30%
30%