Ieee 488.2 commands—description – EXFO FLS-2200 Broadband Source User Manual
Page 120

IEEE 488.2 and Specific Command Reference
112
FLS-2200
IEEE 488.2 Commands—Description
IEEE 488.2 Commands—Description
*CLS
Description
The *CLS command clears the Standard Event
Status Register and the Error/Event Queue.
Syntax
*CLS
Parameter(s)
None
*ESE
Description
The *ESE command sets the Standard Event
Status Enable Register bits, as defined in the
table below. This register contains a mask value
for the bits to be enabled in the Standard Event
Status Register.
Syntax
*ESE<wsp><RegisterValue>
Parameter(s)
RegisterValue:
The program data syntax for <RegisterValue> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
The <RegisterValue>, expressed in base 2,
represents the bit values of the Standard Event
Status Enable Register.
LSB
OPC
N.U.
QYE
N.U
MSB
Standard Event Status Enable Register
DDE
PON
CME
EXE
N.U
EXE
DDE