EXFO FLS-2200 Broadband Source User Manual

Page 150

Advertising
background image

IEEE 488.2 and Specific Command Reference

142

FLS-2200

Product-Specific Commands—Description

:SOURce[1..n]:CURRent[:LEVel]

[:IMMediate][:AMPLitude]?

Description

This query returns the source's current level.

At *RST, the current level is device-dependent.

Syntax

:SOURce[1..n]:CURRent[:LEVel][:IMMediate][:A

MPLitude]?[<wsp>MAXimum|MINimum|DEFa

ult]

Parameter(s)

Parameter 1:
The program data syntax for the first parameter is

defined as a <CHARACTER PROGRAM DATA>

element. The allowed <CHARACTER PROGRAM

DATA> elements for this parameter are:

MAXimum|MINimum|DEFault.

MINimum is used to retrieve the instrument's

smallest supported value.

MAXimum is used to retrieve the instrument's

greatest supported value.

DEFault is used to retrieve the instrument's

default value.

Response Syntax

<Current>

Response(s)

Current:
The response data syntax for <Current> is

defined as a <NR3 NUMERIC RESPONSE DATA>

element.

The <Current> response corresponds to the

current source current drive.