Remote Control Commands
74
FVA-3150
IEEE 488.2 Required Commands
*OPC?
Description This query puts an ASCII 1 in the output queue when the
contents of the input queue has been processed.
Syntax
*OPC?
Response
“1”
Note
The OPC? command is equivalent to the *OPC? command.
Both give the same result.
*RST
Description This command empties the step response list. It is only seen
when it is part of another multiple command. In the example
below, by adding this command after *IDN?, you will not be
able to access the answer. The *RST, in this instance, erases the
identification string. In addition, this command performs the
following operations:
Returns to initial state before command is sent, and not
necessarily to previous settings.
Forces the device to enter into an Operation Complete
Command Idle State (OCIS).
Forces the device to enter into an Operation Complete
Query Active State (OQIS).
Initializes previous responses unless there has been a
program message terminator preceded by an *RST.
Syntax
*RST
Example
*IDN?,*RST<NL>