EXFO FVA-3150 Variable Attenuator User Manual
Page 90

Remote Control Commands
84
FVA-3150
Specific commands-Quick Reference Command Tree
CAL
Description This command stores an offset value that will be applied to all
wavelengths in Absolute, Reference, and Offset modes. This
offset value will be included in the measurement returned by
ATT? but will not be taken into account on the FVA-3150 display.
This offset value is deleted when the FVA-3150 is turned off.
Syntax
CAL<wsp><value>[<wsp><DB>]
Parameters The <value> parameter must be a valid attenuation value in
the format “±99.999”. The units (DB) are optional.
Example
CAL 99.999 DB
CAL?
Description This query returns the offset value that is applied to all
wavelengths in Absolute, Reference, and Offset modes.
Syntax
CAL?
Response
The offset value that is applied to all wavelengths in Absolute,
Reference, and Offset modes in the format “±99.999”. The units
are dB.
Note
This offset value is included in the measurement returned by
ATT? but is not taken into account on the FVA-3150 display.
Example
CAL?