Alarm Message
Check	Vents 
 
 
Low	Battery 
 
 
 
 
XX: Service!*
*Value: 21-50
Description
Device	is	unable	to	maintain 
oxygen	purity.
 
Estimated	battery	life	less	
than	17	minutes. 
 
 
 
Service required.
Action
Be	sure	air	inlet/outlet	
has	not	been	blocked.	If	 
alarm persists, contact 
Customer	Service.
Charge	the	battery	pack	
by plugging in to power 
supply.
Note: The message will be 
automatically cleared when 
plugged in to power supply. 
Contact	Customer	
Service.
High Priority Alarms
Note: These alarms will allow the Portable Oxygen Concentrator to continue operating.
Alarm Message
Check	Cannula 
 
 
 
 
 
 
XX: Service*
*Value: 51-70
Description
No	breath	detected	for	5	
minutes. 
 
 
 
 
 
Service required.
Action 
Check	the	cannula	con-
nection. Be sure to
breathe through nose. If
alarm persists, contact
Customer Service.
Note: The message will be 
automatically cleared when 
breathing is detected. 
Contact Customer
Service.
Medium Priority Alarms