Delphi DL-00183 User Manual
Page 35
 
23
DL-00096 Rev. 7
Immunity test 
 
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electric fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601 test 
level 
±	6kV	contact	
± 8kV air
± 2kV for
power supply
lines
± 1kV for
input/output
lines
±	1kV	
differential mode
± 2kV common
mode
Electromagnetic 
environment/guidance 
Floors	should	be	wood,	
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be	at	least	30%.
Mains	power	quality	
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Guidance and manufacturer’s declaration: electromagnetic immunity
The	Portable	Oxygen	Concentrator	is	intended	for	use	in	the	electromagnetic	 
environment	specified	below.	The	customer	or	the	user	of	the	concentrator	should	
assure	that	it	is	used	in	such	an	environment.
Compliance 
level 
±	15kV	contact	
± 8kV air
± 2kV for
power supply
lines
± 1kV for
input/output
lines
±	1kV	
differential mode
± 2kV common
mode
DL-00096 rev 7.indd 23