Delphi DL-00183 User Manual
Page 30
 
18
DL-00096 Rev. 7
Alarm Message
Replace	C-filtr 
 
 
Replace	P-filtr 
 
 
Wash	I-filtr 
 
 
Replace batt 
 
 
 
 
 
 
 
 
XX: Service*
*Value: 71-99
Description
Compressor	filter	needs	
replacement. 
 
Patient	filter	needs	 
replacement. 
 
Input	filter	needs	to	be	
cleaned. 
 
Main	battery	has	exceeded	
300	recharge	cycles.
 
 
 
 
 
 
Service required.
Action
Replace	filter.	Press	and	
hold mute key for more 
than	4	seconds	to	reset	
the	hour	counter.
Replace	filter.	Press	and	
hold mute key for more 
than	4	seconds	to	reset	
the	hour	counter.
Wash	input	filter.	Press	
and hold mute key for 
more	than	4	seconds	to 
reset	the	hour	counter.
Contact	Customer	
Service	for	replacement.	
Note: This alarm will be 
automatically cleared once 
a new battery is inserted. 
Battery will continue to 
function after 300 recharge 
cycles, but for a shorter time 
than a new battery.
Contact	Customer	
Service.
Low Priority Alarms
Other Messages
Message
Power	removed 
 
SW	Ver	N.NNN 
 
 
Self test 
 
 
Shutting off
External power
Description
External	power	has	been	
disconnected; unit is now 
running	on	battery.
Display	of	software	version	
during startup and calibra-
tion.	This	includes	test	of	
beeper	and	LEDs.
Indicates	unit	is	going	
through	its	test	at	startup.	
Displayed	for	approximately 
10	seconds.
Displayed	while	unit	goes	
through	its	power-down	
sequence.
External	power	is	 
connected.
Action
No	action	is	required. 
 
No	action	is	required. 
 
 
No	action	is	required. 
 
 
No action is required.
No action is required.
DL-00096 rev 7.indd 18
6/1/2009 4:48:34 PM