Appendix 2: Technical Information
ELECTROMAGNETIC COMPATIBILITY (EMC) INFORMATION
Medical	electrical	equipment	requires	special	precautions	regarding	 
electromagnetic	compatibility	(EMC).	Portable	and	mobile	radio	frequency	(RF)	
communications	equipment	can	affect	devices	such	as	the	Portable	Oxygen	
Concentrator.	As	such,	the	device	should	not	be	used	adjacent	to	other	 
equipment.	If	this	is	not	practical,	then	observe	the	device	to	make	sure	it	is	
operating	properly	at	all	times.
Emissions test 
RF	emissions 
CISPR 11
RF emissions
CISPR	11
Harmonic emissions
IEC	61000-3-2
Voltage	fluctuations/ 
flicker emissions
IEC 61000-3-3
Compliance 
Group	1 
Class B
Class A
Complies
Electromagnetic environment/guidance 
The	Portable	Oxygen	Concentrator	uses	
RF energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any interference
in	nearby	electronic	equipment.
The concentrator is suitable for use in all 
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Guidance and manufacturer’s declaration: electromagnetic emissions
The	Portable	Oxygen	Concentrator	is	intended	for	use	in	the	electromagnetic	 
environment	specified	below.	The	customer	or	the	user	of	the	concentrator	should	
assure	that	it	is	used	in	such	an	environment.