Appendix 1: Technical Description
Size: 7.4” W x 4.6” D x 11.6” H (188 mm W x 117 mm D x 295 mm H)
Unit weight: Less than 10 lbs (without carrying case or cart)
Power requirements: 100-240V	AC	(±	10%),	50-60	Hz;	battery	power	18V	DC,	6.7	amps;	DC	adap-
tor,	18V	DC,	6.7	amps	(Note: See Approved Accessories Guide for model and part number of AC 
power supply.)
Purity: 90%	±	3%	(22°C	±	3°C,	14.29	psia	±	0.04	or	739	mmHg	at	40%	±	15%	relative	humidity)	
at all flow rates
Setting: User adjustable 1.0 to 5.0 (with 0.5 increments) (min. 1.0; max. 5.0)
Setting indicator: LCD display
Maximum oxygen discharge pressure: 17 psi at setting of 5.0
Humidity range: 5% to 93% ± 2% non-condensing
Operating altitude: 0’ to 8,000’ relative to sea level (0 km to 2.4 km)
Sound pressure level (measured 1 m from edge of chassis): Less than 50 dB(A) average
Type of protection (electrical): Class I
Degree of protection (electrical): Type BF
Degree of protection (water): IPX2
Degree of safety (flammable anesthetic mixture): Not	suitable	for	use	in	the	presence	of	a	
flammable	anesthetic	mixture
Operation: Continuous	operation	at	temperatures	of	95°F	(35°C)	down	to	41°F	(5°C),	and	short-time	
operation	for	25	minutes	at	temperatures	up	to	104°F	(40°C).	
Alarm sound pressure range: 65 to 85 dB(A)
Alarm system delays: Less	than	10	seconds	after	detection	(low	oxygen	alarms	if	oxygen	is	less	
than	82%	volume	fraction	at	specified	environmental	conditions)
Oxygen concentrator status indicator:	High	priority	alarm	that	indicates	when	oxygen	concentra-
tion	drops	below	82%
Maximum Temperature:		Under	extreme	environmental	and	operating	conditions,	specific	maxi-
mum	surface	temperatures	are	listed	below.	Each	of	these	specific	surfaces	also	lists	an	expected	
user contact limit as follows:
The	Key	panel	can	reach	up	to	140°F	(60°C).	This	surface	is	intended	to	be	touched	for	less	than	one	
minute	at	a	time.		The	clear	plastic	window	on	the	carrying	case	can	reach	up	to	118.4°F	(48°C).	This	
surface	is	intended	to	be	touched	for	less	than	ten	minutes	at	a	time.		The	remainder	of	the	carrying	
case	can	reach	up	to	109.4°F	(43°C).	This	surface	has	no	time	limitation	on	user	contact.