Chapter 3: Getting Started

Product Service 1-800-522-7658

13





NOTE: The drain valve assembly attaches to
the underside of the cabinet bottom.





Apply a light coat of silicone sealant to this
surface of the connector, aligning the three
holes in the connector with the three holes in
the Biosafety Cabinet liner.


Initial Certification

Prior to use, a qualified certifier should certify all Biosafety Cabinets. Under
normal operating conditions, the Biosafety Cabinet should be recertified at
least annually and when relocated or serviced. The certifier should perform
the following tests, as recommended in ANSI/NSF International Standard
Number 49 in effect when the cabinet was manufactured:

• Downflow Velocity Profile Test

• Inflow Velocity Test

• Airflow Smoke Patterns

• HEPA Filter Leak Test

• Vibration Test *

• Noise Level Test *

• Lighting Intensity Test *

*These tests are user comfort related tests and may be omitted at the

user’s or certifier’s discretion.


If you have any questions regarding certification agencies or need assistance
in locating one, contact Labconco’s Product Service Department at 1-800-
522-7658 or 816-333-8811.