Labconco Biological Safety Cabinets 34610 Series User Manual

Page 26

Advertising
background image

Chapter 4: Performance Features & Safety Precautions

Product Service 1-800-522-7658

21


Proper operation of the cabinet depends largely upon its location and the
operator’s work habits. Consult the Installation and Normal Operation
sections of this manual for further details.
Avoid direct exposure of plastic or coated materials to ultraviolet (UV)
radiation. Never bypass the UV safety interlock that only allows the UV
light to work when the sash is closed.

When surface disinfecting the Biosafety Cabinet:

• Avoid splashing the disinfecting solution on skin or clothing.

• Ensure adequate ventilation.

• Carefully follow the disinfectant’s safety instructions.

• Always dispose of disinfecting solutions in accordance with local and

national laws.

• DO NOT allow disinfectants with high concentrations of free chlorine

to contact the stainless steel components of the Biosafety Cabinet for
a long period of time. Free chlorine will corrode stainless steel after
extended contact.

The electrical receptacle cover may be removed to facilitate surface
decontamination. In the event of gross contamination, the cover should
be removed, sterilized and/or decontaminated as required.

Biosafety Cabinets should be decontaminated for any of the following
reasons:

• Before maintenance work requiring entry into contaminated areas.

• Before HEPA filter changes.

• Before performing certification tests requiring entry into

contaminated areas.

• Before relocating the cabinet.

• Before changing research programs.

• After the gross spill of biohazardous material or toxic chemicals.

Advertising
This manual is related to the following products: