Chapter 6: Maintaining The Cabinet

Product Service 1-800-522-7658

47


Changing the Optional UV Lamp:
Note:
For optimum performance, the UV lamp should be changed on an
annual basis.

The UV lamp and the work area of the cabinet must be thoroughly
decontaminated before removing the lamp.

1. Start the cabinet and let it operate for 5 minutes.
2. Raise the sash to its full open position.
3. Thoroughly surface decontaminate the UV lamp and the work area of the

cabinet.

4. Unplug the cabinet or turn off the System Reset Switch, located on the

top of the cabinet.

5. Remove the UV lamp by rotating it 90 degrees and lifting it straight up

and out of its sockets.

6. Install new lamp by reversing the removal procedure.

Resetting a Circuit Breaker:

To reset any of the circuit breakers located on the left side of the electronics
module, depress the white button until it sets.


Circuit Breakers- The
front breaker protects
the electrical outlets,
the rear breaker
protects the motor and
lights.






Figure 6-5