Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

29

Original instructions

The Logic+ Touchpad

The touchpad of the Logic+ is shown in Figure 5-6. Take a moment to get
familiar with the buttons, their locations and functions. Also familiarize
yourself with the display located on the right side wall. The display will
report system functions, such as filter capacity, timer displays, alarm or error
messages, as well as icons that illuminate when cabinet functions such as UV
light and blower are operational.

Blower Button – Starts or stops the cabinet blower. When the
blower is in Smart-Start

TM

mode, opening the sash from the closed

position turns the blower on automatically. Pressing this button
overrides Smart-Start operation.


Light Button – Turns the fluorescent lamps on or off. Closing the
sash automatically turns the lights off. When the lights are in Smart-
Start mode, raising the sash turns the lights on automatically.

Outlet Button – Turns the electrical outlets in the work area on or
off.

UV Light Button – Turns the UV lamp on or off. When the UV
lamp is in Smart-Start mode, closing the sash turns the light on.
When the sash is raised, the light turns off automatically.

Timer Button – Allows you to select either a repeating interval
timer, or an elapsed timer (stopwatch).

OK/Mute Button – Mutes all audible alarms for approximately 5
minutes, unless there is a system error alarm. When in the Menu
mode, this button is used to select an option.

Menu Button – This button toggles the display between the display
and menu modes. When in the menu mode, pressing this button
returns you to the previous menu level.

Select Buttons – Allow you to choose different options in the menu
mode.




Figure 5-6