Chapter 2: Prerequisites

Product Service 1-800-522-7658

3

Original instructions

Location Requirements

Note: The biosafety cabinet should be located away from traffic patterns,
doors, fans, ventilation registers, fume hoods and any other air-handling
devices that could disrupt its airflow patterns. All windows in the room
should remain closed. Figure 2-1 shows the preferred location for the
biosafety cabinet.


Preferred location

Air register blocked or redirected

to prevent cabinet disruption

Alternate location




Main Disconnect Device
(Power cord plug in back
of the electronics module).




Do not position the unit so that it is difficult to operate the main
disconnect device.

Ne placez pas l'appareil de sorte qu'il est difficile de faire fonctionner le
dispositif principal de déconnexion.

Figure 2-1