Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

35

Original instructions

UV Settings

For models equipped with the optional UV light the Logic+ has an integral
UV light maintenance system. It allows you to define how many hours you
want the UV lamp to operate before receiving a reminder to replace it, a way
to monitor how many hours the lamp has been on, and the means to reset the
UV lamp hourmeter.

UV Lamp Hourmeter

This display only shows how many hours the UV lamp has been lit, and how
many hours remain until you will receive a warning to replace the lamp.

From the Settings menu screen, use the

and

buttons on the touchpad to highlight the

UV Settings

option-it will turn white when

selected. Press

OK

to enter the UV Settings

screen. Using the

and

buttons on the

touchpad, highlight UV Hourmeter-it will turn
white when selected, and then press

OK

. The

screen will then return to the UV Hourmeter
screen.

This screen displays how many hours the UV
lamp has operated, and how many hours of
operation remain before replacement is
recommended. When finished with this screen,
press

MENU

to return to the first UV Settings

screen.

Reset UV Lamp Hourmeter

This option lets you reset the UV hourmeter to 0 hours whenever the lamp
has been replaced.

This screen allows you to reset the UV Hourmeter
to 0 hours if you press

OK

. Pressing

MENU

will

return you to the first UV Settings screen without
resetting the hourmeter.