The common service operations necessary to maintain the biosafety cabinet
for peak performance are listed below.
Note: This manual covers operation and maintenance operations for the
owners/users of the Logic+ biosafety cabinets. Complete certification
procedures, service operations and specifications are published in a separate
publication Technical Manual: Purifier

®

Logic

®

+ Biosafety Cabinets. This

manual is available from Labconco’s website:

www.labconco.com

. A

complete certifier service kit is available to qualified certifiers from
Labconco. Call Labconco’s Product Service Department at 800-821-5525 or
816-333-8811.

Do NOT contact blower wheel while still in motion.

NE PAS être en contact avec la roué du ventilateur tant qu’il est
en marche.

Routine Maintenance Schedule

Weekly

• Wipe down the interior surfaces of the cabinet with a suitable

disinfectant, or 70% ethanol, and allow to dry.

• Using an appropriate glass cleaner, or LabSolutions

TM

Glass &

Surface Wipes, Labconco part # 1570000; clean the sash and the
surface of the UV lamp, if so equipped.

• Operate the cabinet blower, noting the percent filter life remaining in

an operational log.

Monthly (or more often as required)

• Using a damp cloth, or LabSolutions Glass & Surface Wipes,

Labconco part # 1570000; clean the exterior surfaces of the cabinet,