Chapter 6: Maintaining The Cabinet

Product Service 1-800-522-7658

43

Original instructions

Front Grille Removal:

Note: The grille must be thoroughly decontaminated before removing it.

1. Remove the work surface as described earlier.

2. At one end of the grille, grip the front of grille with one hand, and the

back with the other hand. Pivot that end of the grille upward and
inward, paralleling the angle of the sash, as shown in Figure 6-1.

3. Pull the other end of the grille up and away from the bottom edge of

the cabinet.

4. Reinstall the grille by reversing the above sequence, ensuring that the

grille properly engages the bottom edge of the cabinet.

Towel Catch Removal:

Although not normally required, the towel catch can be removed for cleaning,
inspection, etc.

Note: The work surface of the cabinet and the towel catch must be
thoroughly decontaminated before removing either.

1. Remove the work surface as described above.

2. Remove the towel catch by pivoting the bottom out toward you, as shown

in Figure 6-2; it is spring loaded, and you will feel some resistance. Note
the orientation of the towel catch.

3. Surface decontaminate the towel catch before removing it.

4. Reinstall the towel catch by sliding it back into position, in the correct

orientation. Also ensure that the Sampling/Decontamination Tube (the

Figure 6-1

Advertising