Labconco Protector Demonstration Hoods 3945021 User Manual

Page 57

Advertising
background image

Product Service 1-800-522-7658

53



A

A

P

P

P

P

E

E

N

N

D

D

I

I

X

X

D

D

R

R

E

E

F

F

E

E

R

R

E

E

N

N

C

C

E

E

S

S

Many excellent reference texts and booklets are currently available. The
following is a brief listing:

Laboratory Ventilation Standards

Federal Register 29 CFR Part 1910
Non-mandatory recommendations from “Prudent Practices”.
• Fume hoods should have a continuous monitoring device

• Face velocities should be between 60-100 linear feet per minute (lfpm)

• Average 2.5 linear feet of hood space per person

Occupational Health and Safety
U.S. Department of Labor
200 Constitution Avenue N.W.
Washington, DC 20210
(202) 523-1452
www.osha.gov


Industrial Ventilation-ACGIH
• Fume hood face velocities between 60-100 lfpm

• Maximum of 125 lfpm for radioisotope hoods

• Duct velocities of 1000-2000 fpm for vapors, gasses and smoke

• Stack discharge height 1.3-2.0 x building height

• Well designed fume hood containment loss, <0.10 ppm

Industrial Ventilation, A Manual of Recommended Practice.
24

th

Edition, 2001

American Conference of Governmental Industrial Hygienists
1330 Kemper Meadow drive
Cincinnati, OH 45240-1634
(513) 742-2020
www.acgih.org

Advertising