Chapter 3: Getting Started

Product Service 1-800-522-7658

11

Solvent Safety Precautions

CAUTION: The CentriVap micro IR is not classified as “explosion proof.”
It has been designed with safety as a primary

consideration

and should be

used in a prudent manner using “good laboratory practices.” It has been
designed for use with compounds as described in the United States National
Electrical Code Class I, Group D. The heater may be programmed to run
as hot as 65°C, however, the heater elements may be hotter. An over
temperature protection device limits the heater to a maximum temperature
of 80°C. It is important that the solvents used are compatible with these
temperatures. Do not evaporate solvents that have an autoignition
temperature below 180°C. Do not evaporate solvents that are classified as
Group A, B, or C by the National Electrical Code. Evaporate only non-
flammable or Group D solvents with autoignition temperatures 180°C or
above. Use of other compounds could cause an explosion.

CAUTION: Solvents used in the CentriVap micro IR may be flammable or
hazardous. Use extreme caution and keep sources of ignition away from the
solvents. When using flammable or hazardous solvents, the CentriVap micro
IR should be operated inside a fume hood.
If a sample is spilled in the chamber it must immediately be cleaned up.
Hazardous materials such as strong acids or bases, radioactive substances
and volatile organics, must be handled carefully and promptly cleaned up if
spilled.
Do not store flammable or hazardous solvents within 12 inches (300 mm) of
the CentriVap micro IR.

IMPORTANT: The disposal of substances used in connection with this
equipment may be governed by various Federal, State or local regulations.
All users of this equipment are urged to become familiar with any
regulations that apply in the user’s area concerning the dumping of waste
materials in or upon water, land or air and to comply with such regulations.

Advertising