Chapter 5: Maintaining Your CentriVap micro IR

Product Service 1-800-522-7658

18

7. Check the continuity of the protective earth ground between the ground

terminal on the power inlet and a bare metal housing panel. Contact
Labconco if there is no continuity.

8. Repair any defects to the surface where the unit is installed.

Monthly:

1. The rubber and silicone components on the CentriVap may eventually

deteriorate and require replacement. The effective life of those parts depends
upon both their usage and the surrounding environment. Check all hoses and
gaskets and replace any that show signs of hardening, permanent set or
deterioration.

2. Using a soft cloth, sponge or chamois and a mild, non-abrasive soap or

detergent, clean the glass lid.

3. Using a soft cloth, sponge, or chamois and a mild, non-abrasive soap or

detergent, clean the exterior surfaces of the unit. Liquid spray cleaners and
polishes may be used on the exterior surfaces. Do not use solvents to remove
stains from the exterior surfaces as they may damage the finish.