Before you install your CentriVap micro IR, you need to prepare your site for
installation. Carefully examine the location where you intend to install your
CentriVap. You must be certain that the area is level and of solid construction. In
addition, an exhaust means must be provided. An electrical source must also be
located near the installation site.

Carefully read this chapter to learn:

• The electrical supply requirements.

• The exhaust requirements.

• Space requirements.

Refer to Appendix C: CentriVap micro IR Specifications for complete CentriVap
electrical and environmental conditions, specifications and requirements.

Electrical Requirements

The CentriVap micro IR Concentrator requires a grounded electrical outlet. This
outlet requires a 15 Amp circuit breaker or fuse for models rated at 115V (60 Hz).
An 8 Amp circuit breaker or fuse is required for models rated at 230V (50/60 Hz).
Various models are available for use in different locations throughout the world.
Each has a unique plug suitable for the specific location. See CentriVap
Components in Chapter 3 for detail plug specifications. If the plug supplied with
the CentriVap Does not match the available receptacle, remove this plug and
replace it with an approved plug of the suitable style.

It is recommended that an emergency switch for disconnecting the power in the
case of a malfunction be located remote from the CentriVap micro IR, preferably
outside the room in which the unit is housed, or adjacent to the exit from that
room.