Refer to the following if your CentriVap micro IR fails to operate properly. If the
suggested corrective actions do not solve your problem, contact Labconco for
additional assistance.
CAUTION: Disconnect power before corrective action is taken.


PROBLEM

CAUSE

CORRECTIVE ACTION

Unit will not operate

Unit not connected to
electrical power

Connect unit to proper electrical receptacle.

Circuit breaker blown

Correct electrical problem and reset circuit
breaker.

Lid open

Close lid.

Excessive vibration

Sample vials not located
symmetrically in rotor

Reposition sample vials.

Sample odor in lab

Vent hose exhausting into
lab area

Redirect hose to fume hood.

Evaporation rate is
reduced

Heater inoperable

Contact Labconco.

Vacuum pump failure

Check pump.

Liquid in vacuum line

Clear vacuum line