Chapter 6: Using the RS232 Receptacle

Product Service Domestic 1-800-522-7658, International 816-333-8811

1. Data Rate 2400 Baud
2. 8 Bit word length
3. 1 Start bit, 1 Stop bit
4. No parity is transmitted
5. Standard ASCII character set


The time between data transmissions may be varied by the user to
occur at 10, 30, 60, 300 or 600 second intervals. Press the MENU
button until the RS232 screen appears on the display.






Press select until the desired time interval is shown.

The format of the transmitted message is as follows:

B<T= -48 V= 18 SF = 999 C=999 PG=9 >


Where:

B = Base Unit

T = Collector Temperature

-48 = Collector Temperature in °Celsius

V = Vacuum

18 = Vacuum Level in Micro Bars

SF = Shell Freezer Temperature (if so

equipped)

9 = Feature not installed or out of range

C = Vacuum Chamber Temperature (if so

equipped)

PG = Purge (if so equipped)

1 = On
0 = Off


If a Stoppering Tray Dryer is attached to the Freeze Dryer and the
data interconnect cable is installed, additional data about that
accessory will be transmitted from the Freeze Dryer. See the
User’s Manual for the Stoppering Tray Dryer for a description of
the data.